Trends in Maintenance Status and Usability of Public Automated External Defibrillators During a 5-Year On-Site Inspection
Kim TY, Jung Y, et al. · Scientific Reports · Sci Rep. 2022;12:10738
Inspectors checked public AEDs in Seoul, South Korea against a standard checklist every year from 2013 to 2017, a total of 23,619 inspections.
Key Findings
- 15.4% of the AEDs inspected were not ready to use.
- Invalid (expired) electrode pads were the leading cause, involved in 83.7% of not-ready AEDs, ahead of device malfunction (22.1%) and poor battery status (17.2%).
- The share of AEDs with valid pads fell from 90.0% to 87.2% over five years, even as access, signage and device operation improved.
- 44.1% of AEDs were not usable around the clock, most often because the building was not accessible 24 hours a day.