AED White Papers & Research

What independent research and expert panels have found about AED readiness, maintenance, accessibility, public policy and survival, summarized in plain language with links to every original source.

At a Glance

What the Research Shows

Across countries and study designs, the same pattern appears: AEDs that look ready often are not, and expired pads, weak batteries and restricted access are the usual reasons.

15.4%of public AEDs inspected in Seoul were not ready to useKim et al., Scientific Reports, 2022
21%of U.S. public AEDs tested failed at least one phase of testingSutton et al., AHA Scientific Sessions, 2017
83.7%of not-ready AEDs had invalid (expired) electrode padsKim et al., Scientific Reports, 2022
~4%of U.S. out-of-hospital cardiac arrests involve an AEDCabañas et al., JAHA, 2025

Research Library

White Papers & Studies

Peer-Reviewed StudyReadiness & Maintenance2022

Trends in Maintenance Status and Usability of Public Automated External Defibrillators During a 5-Year On-Site Inspection

Kim TY, Jung Y, et al. · Scientific Reports · Sci Rep. 2022;12:10738

Inspectors checked public AEDs in Seoul, South Korea against a standard checklist every year from 2013 to 2017, a total of 23,619 inspections.

Key Findings

  • 15.4% of the AEDs inspected were not ready to use.
  • Invalid (expired) electrode pads were the leading cause, involved in 83.7% of not-ready AEDs, ahead of device malfunction (22.1%) and poor battery status (17.2%).
  • The share of AEDs with valid pads fell from 90.0% to 87.2% over five years, even as access, signage and device operation improved.
  • 44.1% of AEDs were not usable around the clock, most often because the building was not accessible 24 hours a day.
Research PresentationReadiness & Maintenance2017

Readiness of Public Access Defibrillators

Sutton B, Heimroth J, Crawford S, Sutton E, Matzke J · American Heart Association Scientific Sessions; reported by UofL News (April 21, 2017) · University of Louisville News, April 21, 2017

Researchers tested 322 public, non-hospital AEDs at 190 sites in four U.S. regions (Seattle, Suffolk County NY, Central Illinois and Louisville), following each manufacturer’s testing guidelines.

Key Findings

  • 21% of the AEDs failed at least one phase of testing.
  • 5% had expired batteries and would not power on.
  • In regions where more than 80% of AEDs were registered, battery failures were rare (none in Seattle, 2% in Suffolk County).
  • In regions with under 25% registration, test failure rates reached 19.8% (Louisville) and 38.2% (Central Illinois).
Expert ConsensusPolicy & Standards2025

American Heart Association Automated External Defibrillator Symposium: Summary and Recommendations

Cabañas JG, et al. · Journal of the American Heart Association · J Am Heart Assoc. 2025;14(7):e039291

A summary of the November 2023 AHA AED Symposium, where scientists, clinicians, industry and regulators reviewed why AEDs remain underused and what should change.

Key Findings

  • AEDs are used in only about 4% of out-of-hospital cardiac arrests in the United States.
  • The panel found no clear, universal standards for AED maintenance and upkeep, and called for standardized, simple maintenance and readiness plans.
  • It suggested modeling AED readiness on fire-extinguisher programs with regular inspection and testing, and Wi-Fi notifications for expiring batteries and pads.
  • Recommendations also include 24/7-accessible placement, local AED registration and universal AED signage.
Scientific ReviewAccess & Placement2024

The Impact of Locked Cabinets for Automated External Defibrillators (AEDs) on Cardiac Arrest and AED Outcomes: A Scoping Review

Oonyu L, Perkins GD, Smith CM, Vaillancourt C, Olasveengen TM, Bray JE; ILCOR BLS Task Force · Resuscitation Plus · Resusc Plus. 2024;20:100791

An international review of 10 studies examining whether locking AED cabinets prevents theft and vandalism, and how locks affect access in an emergency.

Key Findings

  • Theft, vandalism and missing-AED rates were generally below 2%, including at sites accessible 24/7.
  • Simulation studies found that locked or staff-controlled cabinets slowed AED retrieval.
  • None of the included studies measured patient outcomes.
  • The 2022 ILCOR guidance referenced in the review advises against locking AED cabinets.
White PaperPolicy & Standards2016

Legislative Strategies for Modernizing U.S. AED Laws

Lazar RA · Readiness Systems, LLC · Readiness Systems white paper, March 2016

A review of how state AED Good Samaritan laws were written as of March 2016, and of the gaps that may discourage organizations from deploying AEDs.

Key Findings

  • By the author’s criteria, only 5 states (10%) offered meaningful Good Samaritan immunity for AED programs at the time.
  • 34 states (67%) appeared to limit immunity to activities during AED use, leaving program decisions such as placement and training unprotected.
  • 19 states (37%) appeared to make immunity conditional on meeting operational requirements such as maintenance, training or agency notification.
  • The author concludes that well-run AED programs matter more than the laws themselves.
Peer-Reviewed StudyClinical Evidence2025

Escalating vs Fixed Energy Defibrillation in Out-of-Hospital Cardiac Arrest Ventricular Fibrillation

Tang H, Wu R, Yin L, et al. · JAMA Network Open · JAMA Netw Open. 2025;8(4):e257411

A cohort study of 342 patients in 48 cities across China (2017–2023), comparing AEDs programmed with escalating higher-energy shocks against AEDs using a fixed lower energy.

Key Findings

  • The first shock briefly stopped ventricular fibrillation in about 93–94% of cases in both groups.
  • Only about half stayed terminated until the next rhythm analysis (49% vs 47%).
  • Escalating higher-energy regimens more often restored a sustained organized rhythm (34% vs 25%).
  • The difference was greatest in refractory cases after the third shock (35% vs 18%).
Scientific ReviewClinical Evidence2001

Improving Survival From Sudden Cardiac Arrest: The Role of the Automated External Defibrillator

Marenco JP, Wang PJ, Link MS, Homoud MK, Estes NAM 3rd · JAMA · JAMA. 2001;285(9):1193–1200

A foundational review of 101 peer-reviewed articles on AED technology, time to defibrillation and survival from sudden cardiac arrest.

Key Findings

  • Advances in AED design simplified operation, improved accuracy and reduced cost.
  • AED use by first responders and laypeople shortened time to defibrillation and improved survival in several communities.
  • Use by both traditional and nontraditional responders appeared safe and effective.
  • Early cost-effectiveness studies of AED use were favorable.